AN ANALYSIS OF REPEATED EXAMINATIONS IN CONVENTIONAL FILM–SCREEN RADIOGRAPHY (FSR).
Main Article Content
Abstract
Background: X-ray is an ionising form of radiation used in conventional radiography and this can result in deleterious biological changes in the body if not regulated. Radiation safety is an important practice in a medical facility that utilizes radiation in any form for diagnosis, intervention or treatment. Repeat of non-diagnostic radiographs add to the radiation received by the patient and the personnel. The committee on quality assurance in diagnostic X-ray recommended a repeat rate of 5 – 7% or less.
Materials and Methods: It was a prospective study involving all routine radiographic cases in department over a two month period during which 1251 examinations were carried out. Radiographs were assessed for diagnostic quality by Radiologists following the usual clinical protocols in radiology film review. The number of radiographs in each examination type which did not meet diagnostic criteria and therefore required repeat, were isolated and counted. Results are presented in simple percentages.
Results: The overall repeat rate was 8.6%. The highest repeat rate was observed in radiographs of the lumbosacral region (53.06%) and the lowest rate in the leg (2%). Under-exposure was the leading cause of repeat (41.67%), while film fog was the least cause (0.93%). Inappropriate selection of exposure factors and poor patient positioning by radiographers who are mainly interns (n=8) and junior radiographers (n=4) contributed most to the repeats. Patient faults or poor darkroom practice were the other reasons for repeats.
Conclusion: The rate of repeat is above the accepted limits recommended by the committee on quality assurance in diagnostic X-ray (5 – 7% or less). It has increased above the 4% previously obtained (1992 assessment). The rate can be reduced if the junior cadre radiographers are better supervised and a quality assurance program instituted.
Downloads
Article Details
Section

This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
All articles in JRRS are published under the Creative Commons Attribution 4.0 International License (CC BY 4.0). This permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
How to Cite
References
Honea R, Bladop ME, Ma Y. Is reject analysis necessary after converting to computed radiography? Journal of Digital Imaging, 2002; 5(Suppl 1). ISSN: 0892-1889.
Shalemaei RR. Film reject analysis for conventional radiography in Iranian main hospitals. Radiation Protection Dosimetry, 2011 Sep; 147(1-2):220–222. Epub 2011 Jul 14.
Committee on Quality Assurance in Diagnostic X-ray. Quality control recommendations for diagnostic radiography. Volume 2: Podiatric facilities. Publication 01–5, July 2001.
Bassey DE. Analysis of wasted X-ray films in the University of Calabar Teaching Hospital. West African Journal of Radiology, 1997 Apr; 4(1):28–32.
Peer S, Peer R, Giacomuzzi SM, Jaschke W. Comparative reject analysis in conventional film-screen and digital storage phosphor radiography. Radiation Protection Dosimetry, 2001; 94(1-2):69–71.
Lewentat G, Bohndorf K. Analysis of reject X-ray films as a quality assurance element in diagnostic radiology. Rofo, 1997 May; 166(5):376–381.
Eze KC, Omodia N, Okegbunam B, Adewoniyi T, Nzotta CC. An audit of rejected/repeated X-ray films as a quality assurance element in a radiology department. Nigerian Journal of Clinical Practice, 2008 Dec; 11(4):355–358.
Waaler D, Hofmann B. Image rejects/retakes—Radiographic challenges. Radiation Protection Dosimetry, 2010 Apr–May; 139(1-3):375–379. Epub 2010 Feb 16.
Al-Malki MA, Abulfaraj WH, Bhuiyan SI, Kinsara AA. A study on radiographic repeat rate data of several hospitals in Jeddah. Radiation Protection Dosimetry, 2003; 103(4):323–330.
Akhtar W, Aslam M, Ali A, Mirza K, Ahmed N. Film retakes in digital and conventional radiography. Journal of the College of Physicians and Surgeons Pakistan, 2008 Mar; 18(3):151–153.
Bassey CE, Ojo OO, Akpabio I. Repeat profile analysis in an X-ray Department. Journal of Radiological Protection, 1991; 11(3):179–183.