X-RAY FILM REJECT ANALYSIS AS A QUALITY INDICATOR IN A TERTIARY HEALTH CENTRE IN NORTHWESTERN NIGERIA

Main Article Content

Sadiq Abubakar Audu
Miftaudeen Mutiat Nike
Ibrahim Rumaysa
Mohammed Abacha
Umar Abubakar
Muftaudeen Basirat

Abstract

Objective: To assess the rate of rejects in routine radiography as a quality control measure.


Methods: Retrospective study was carried out on radiographs obtained from two conventional diagnostic rooms of the Radiology Department at Usmanu Danfodiyo University Teaching Hospital, Sokoto, for a period of six months (July-December 2016). All rejected radiographs were analyzed under uniform viewing conditions. All images with good diagnostic quality, those carried out for special examination and mammograms were excluded. Data were collected and entered into database for analysis.


Results: The overall reject rate was found to be 16.4% with chest contributing the (7.1%). The major cause of reject was inappropriate collimation (18.1%).


Conclusion: A total of 16.4% reject was noted in the study conducted, which was above the recommended level given by World Health Organization (WHO).

Downloads

Download data is not yet available.

Article Details

Section

Articles

How to Cite

Audu, S. A., Nike, M. M., Rumaysa, I., Abacha, M., Abubakar, U., & Basirat, M. (2026). X-RAY FILM REJECT ANALYSIS AS A QUALITY INDICATOR IN A TERTIARY HEALTH CENTRE IN NORTHWESTERN NIGERIA. Journal of Radiography and Radiation Sciences, 31(1), 38-44. https://doi.org/10.82547/jrrs.vol31no1.90

References

Owusu-Banahene, J., Darko, E. O., Hasford, F., Addison, E. K., & Okyere, A. J. (2014). Film reject analysis and image quality in diagnostic radiology department of a teaching hospital in Ghana. Journal of Radiation Research and Applied Sciences. https://doi.org/10.1016/j.jrras.2014.09.012

International Commission on Radiological Protection. (1990). Recommendation of the International Commission on Radiological Protection. Oxford, UK: Pergamon Press.

Geijer, H., Beckman, K. W., Andersson, T., & Persliden, J. (2001). Radiation dose and image quality with a flat panel amorphous silicon digital detector. European Radiology, 11(2 Suppl. 1), S280.

Almen, A., Loof, M., & Mattsson, S. (1996). Examination technique, image quality and patient dose in pediatric radiology: A survey including 19 Swedish hospitals. Acta Radiologica, 337–342.

Saia, D. A. (2009). Radiography prep (5th ed., pp. 231–252). New York, NY: McGraw-Hill.

Jabbari, N., Zeinali, A., & Rahmatnezhad, L. (2012). Patient dose from radiographic rejects/repeats in radiology centers of Urmia University of Medical Sciences, Iran. Health, 4, 94–100. http://dx.doi.org/10.4236/health.2012.42015

Yousef, M., Edward, C., Ahmed, H., Bushara, L., Hamdan, A., & Elnaiem, N. (2013). Film reject analysis for conventional radiography in Khartoum hospitals. Asian Journal of Medical Radiological Research, 1(1).

World Health Organization. (2016). Cancer mortality and morbidity. Global Health Observatory (GHO) Data. www.who.int/entity/gho/en

Jemal, A., Bray, F., Ferlay, J., Forman, D., O'Brien, M., Center, M., … & Parkin, D. M. (2011). Cancer burden in Africa and opportunities for prevention. Cancer. www.wileyonlinelibrary.com

Abdulkareem, F. (2009). Epidemiology and incidence of common cancers in Nigeria. Cancer Registration and Epidemiology Workshop.

Vaiserman, A. M. (2010). Radiation hormesis: Historical perspective and implications for low-dose cancer risk assessment. Dose-Response, 8, 172–191.

Ofori, K. E., Antwi, K. W., Arthur, W., Yeboah, A. C., & Tetty, D. K. (2013). Analysis and economic implications of X-ray film repeat/reject in selected hospitals in Ghana. West African Journal of Radiology, 20(1), 14–18.

Conference of Radiation Control Program Directors (CRCPD). (2009). Quality assurance collectible (QA Collectible), sponsored by CRCPD committee on quality assurance in diagnostic X-ray (H-7), repeat analysis.

Sadiq, A. A., Miftaudeen, M. N., Mohammed, A., Akpaniwo, G. M., Girei, A., Garba, I., & Nwobi, I. C. (2017). Reject-repeat analysis of plain radiographs as a quality indicator at University of Maiduguri Teaching Hospital (UMTH). European Journal of Pharmaceutical and Medical Research, 4(2), 188–191.

Nwobi, I. C., Agwu, K. K., Garba, I., & James, P. (2011). Analysis of rejected films at the University of Maiduguri Teaching Hospital (UMTH). Nigerian Journal of Medical Imaging and Radiation Therapy, 1(2), 23–26.

Osahon, O. D., Asemota, O. I., Olajide, B. O., & Igbinovia, O. J. (2016). Analysis of rejected films of selected examination in a tertiary radio-diagnostic center in Benin City, Edo State, Nigeria. Palgo Journal of Medical and Medical Sciences, 3(1), 38–41. http://www.palgojournals.org/PJMMS/Index.htm

Teferi, S., Zewdneh, D., Admassie, D., Nigatu, B., & Kebeta, K. (2012). X-ray film reject rate analysis at eight selected governmental hospitals in Addis Ababa, Ethiopia, 2010. Ethiopian Journal of Health Development, 26(1), 54–59. www.ajol.info

Eze, K. C., Omodia, N., Okegbunam, B., Adewonyi, T., & Nzotta, C. C. (2008). An audit of rejected repeated X-ray films as a quality assurance element in a radiology department. Nigerian Journal of Clinical Practice, 11(4), 355–358. www.ncbi.nlm.nih.gov

Abubakar, A., Muhammad, I. G., Gunda, N. M., Nwobi, I. C., Yusuf, M., Luntsi, G., Muhammad, G. M., & Shettima, A. B. (2016). Assessment of X-ray film reject rate in two tertiary healthcare institutions in Northeastern Nigeria. The International Journal of Science & Technoledge, 4(8), 54–56. www.theijst.com

World Health Organization (WHO). (1980). Workshop on quality assurance in diagnostic radiology, Geneva, October 20–24, 1980.

Most read articles by the same author(s)

Similar Articles

You may also start an advanced similarity search for this article.