DEVELOPMENT AND OPTIMISATION OF SURFACE AND VOLUME RADIOFREQUENCY COILS SUITABLE FOR FAST-FIELD-CYCLING MAGNETIC RESONANCE IMAGING (FFC-MRI)
Main Article Content
Abstract
Objective: To evaluate a modelling and design methodology employed in constructing and optimizing radiofrequency (RF) coils suitable for use with the whole-body fast-field-cycling Magnetic Resonance Imaging (FFC-MRI). It is also aimed at comparing the sensitivity and the signal-to-noise ratio (SNR) of the various types of surface RF coils constructed at the initial and final stages of this research.
Methodology: An experimental study carried out at Biomedical MRI Laboratory at University of Aberdeen. Various designs of RF coil were constructed, optimized and tested with network/signal analyser for use with an experimental FFC - MRI scanner, operating at a detection magnetic field of 0.2 T (proton Larmor frequency of 8.5 MHz). The coils comprised circular loop (CL) RF – receive surface coil and a birdcage RF – transmit volume coil. The intrinsic parameters of the CL coils were measured using a search-coil field probe and a network/signal analyser.
Results: The CL surface coil constructed with copper wire had resonant frequency of 8.46 MHz and Quality factor (Q – factor) of 47.1 while the resonant frequency of the one constructed with litz wire was 8.54 MHz and Q - factor of 85.4 MHz. The intrinsic parameter of the birdcage volume coil was 8.48 MHz and Q - factor of 102.
Conclusion: Bench testing of the coils showed promise as receiver and transmit coils for the FFC-MRI system.
Downloads
Article Details
Section

This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
All articles in JRRS are published under the Creative Commons Attribution 4.0 International License (CC BY 4.0). This permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
How to Cite
References
Abragam A. The Principles of Nuclear Magnetism. Oxford University Press, Oxford, 1961; pp. 4–5.
Brunner DO, Zanche ND, Fröhlich J, Paska J, Pruessmann KP. Travelling-wave nuclear magnetic resonance. Nature, 2009; 457(1): 994–998.
Ferrante G, Sykora S. Technical aspects of fast field cycling. Advances in Inorganic Chemistry, 2006; 57(3): 405–470.
Lurie DJ, Bussell DM, Bell LH, Mallard JR. Proton-electron double magnetic resonance imaging of free radical solutions. Journal of Magnetic Resonance, 1988; 76(2): 366–370.
Lurie DJ, Broche LM, Choi CH, Davies GR, Ismail SR, O Hogain D, Pine KJ. Fast field-cycling magnetic resonance imaging. AMPERE NMR School, Wierzba, Poland, 2010; p. 37.
Lurie DJ, Broche LM, Davies GR, Payne NR, Ross PJ, Zampetoulas V. Field-cycling MRI – a curiosity or the next big thing? AMPERE NMR School, Zakopane, Poland, 2016; p. 11.
Lurie DJ, Aime S, Baroni S, Booth NA, Broche LM, Choi C, Davies GR, Ismail SR, O Hogain D, Pine KJ. Fast field-cycling magnetic resonance imaging. Comptes Rendus Physique, 2010; 11(1): 136–138.
Woo DC, Choe BY, Ha SH, Choi CB. Evaluation of the turn variation of spiral RF surface coils for microscopic imaging and spectroscopy. Measurement, 2007; 40(3): 615–622.
Doyle VL, Payne GS, Collins DJ, Verrill MW, Leach MO. Quantification of phosphorus metabolites in human calf muscle and soft-tissue tumours from localized MR spectra acquired using surface coils. Physics in Medicine and Biology, 1997; 42(1): 691–706.
Meyerspeer M, Krssak M, Kemp GJ, Roden M, Moser E. Dynamic interleaved ¹H/³¹P STEAM MRS at 3 Tesla using a pneumatic force controlled plantar flexion exercise rig. Magnetic Resonance Materials in Physics, Biology and Medicine, 2005; 18(2): 257–262.
Alecci M, Romanzetti S, Kaffanke J, Celik A, Wegener HP, Shah NJ. Practical design of a 4 Tesla double-tuned RF surface coil for interleaved ¹H and ²³Na MRI of rat head. Journal of Magnetic Resonance, 2006; 181(3): 203–211.
Roemer PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM. The NMR phased array. Magnetic Resonance in Medicine, 1990; 16(2): 192–225.
Wright SM, Wald LL. Theory and application of array coils in MR spectroscopy. NMR in Biomedicine, 1997; 10(4): 394–410.
Weiger M, Pruessmann KP, Leussler C, Roschmann P, Boesiger P. Specific coil design for SENSE: a six-element cardiac array. Magnetic Resonance in Medicine, 2001; 45(2): 495–504.
Sodickson DK, McKenzie CA, Ohliger MA, Yeh EN, Price MD. Recent advances in image reconstruction, coil sensitivity calibration, and coil array design for SMASH and generalized parallel MRI. Magnetic Resonance Materials in Physics, Biology and Medicine, 2002; 13(1): 158–163.
McRobbie DW, Moore EA, Graves MJ, Prince MR. MRI from Picture to Proton. Cambridge: Cambridge University Press, 2003; pp. 175–183.
Anoardo E, Galli G, Ferrante G. Fast-field-cycling NMR: applications and instrumentation. Applied Magnetic Resonance, 2001; 20(2): 365–404.
Alfonsetti M, Mazza T, Alecci M. Optimisation of multi-element transverse field radiofrequency surface coils. Measurement Science and Technology, 2006; 17(1): 53–59.
Hutchison JMS. Frequency and power considerations for in-vivo EPR and related techniques. In: Berliner LJ, ed. In-vivo EPR (ESR): Theory and Applications. Springer, USA: Biological Magnetic Resonance, 2003; 18(3): 41–59.
Sullivan CR, Charles R. Optimal choice for number of strands in a Litz-wire transformer winding. IEEE Transactions on Power Electronics, 1999; 14(2): 283–291.