COMPUTED TOMOGRAPHY DOSE FOR ADULT HEAD SCAN IN ANAMBRA STATE OF NIGERIA
Main Article Content
Abstract
Background: Computed tomography is associated with relatively high radiation doses and could cause serious health risks. Globally, it is reported that many physicians do not to have adequate knowledge about CT dose. Furthermore, although dose records are available from developed countries, there is a paucity of literature in Nigeria. Dose outputs in our locality are also scarce in the literature.
Objective: To review adult head CT dose in the four largest centres in Anambra State.
Methodology: A 6-month retrospective retrieval of dose summary from the control console. Digital folders of subjects ≥ 18 years were sampled purposively and sequentially, out of the 2015 population of CT examinations. The CTDIvol and DLP for each case were recorded in a pro forma. The mean dose in each centre as well as the combined mean for all centres were calculated. The results were compared with the recommendations of the European Commission and similar studies from Nigeria.
Results: The digital folders of 200 subjects made up of 104 (52.0 %) males and 96 (48.0%) females with an age range of 18 – 93 years were involved in the study. Cranium (n = 164; 82 %) dominated the CT requests. The mean CTDIvol and DLP in the four centres was 58 mGy and 1112 mGy.cm. The mean CTDIvol (73 mGy) and DLP (1613 mGy-cm) in one of the centres was extreme. When excluded, the mean CTDIvol and DLP for the remaining three centres were 52 mGy and 945 mGy.cm, respectively.
Conclusion: Dose output in Anambra State was comparable to the recommendation of the European Commission but varied significantly from other local studies. The establishment of diagnostic reference levels for CT procedure in the locality is imperative.
Downloads
Article Details
Section

This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
All articles in JRRS are published under the Creative Commons Attribution 4.0 International License (CC BY 4.0). This permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
How to Cite
References
Kalra, M. K., Maher, M. M., Toth, T. L., Hamberg, L. M., Blake, M. A., Shepard, J. A., et al. (2004). Strategies for CT radiation dose optimization. Radiology, 230, 619–628.
Foley, S. J., McEntee, M. F., & Rainford, L. A. (2012). Establishment of CT diagnostic reference levels in Ireland. British Journal of Radiology, 85, 1390–1397.
Wall, B. F., & Shrimpton, P. C. (1998). The historical development of reference doses in diagnostic radiology. Radiation Protection Dosimetry, 80, 15–19.
Hall, E. J., & Brenner, D. J. (2008). Cancer risks from diagnostic radiology. British Journal of Radiology, 81, 362–785.
Neumann, R. D., & Bluemke, D. A. (2010). Tracking radiation exposure from diagnostic imaging devices at the NIH. Journal of the American College of Radiology, 7, 87–89.
Rehani, M., & Frush, D. (2010). Tracking radiation exposure of patients. Lancet, 376, 754–755.
McNitt-Gray, M. F. (2002). AAPM/RSNA physics tutorial for residents: Topics in CT—Radiation dose in CT. RadioGraphics, 22, 1541–1553.
International Electrotechnical Commission. (2002). Medical electrical equipment—Part 2–44: Particular requirements for the safety of X-ray equipment for computed tomography. Geneva, Switzerland: IEC.
McCollough, C. H., Leng, S., Yu, L., Cody, D. D., Boone, J. M., & McNitt-Gray, M. F. (2011). CT dose index and patient dose: They are not the same thing. Radiology, 259, 311–316.
American College of Radiology. (2007). White paper on radiation dose in medicine. Journal of the American College of Radiology, 4, 272–284.
Lee, W. J., Woo, S. H., Seol, S. H., Kim, D. H., Wee, J. H., Choi, S. P., Jeong, W. J., Oh, S. H., Kyong, Y. Y., & Kim, S. W. (2016). Physician and nurse knowledge about patient radiation exposure in the emergency department. Nigerian Journal of Clinical Practice, 19, 502–507.
Famurewa, O. C., Ayoola, O. O., Ogunsemoyin, A. O., & Onyade, A. A. (2013). Radiation protection and dose awareness among doctors in a Nigerian teaching hospital: A preliminary study. West African Journal of Radiology, 20, 37–40.
Muhogora, W. E., & Rehani, M. M. (2009). Patient doses in CT and radiography in Africa. IFMBE Proceedings, 25, 536–539.
Ogbole, G., & Obed, R. (2014). Radiation doses in computed tomography: Need for optimization and application of dose reference levels in Nigeria. West African Journal of Radiology, 21, 1–6.
Garba, I., Engel-Hills, P., Davidson, F., & Tabari, A. M. (2015). Computed tomography dose index for head CT in Northern Nigeria. Radiation Protection Dosimetry, 165, 98–101.
Abdullahi, M., Shittu, H., Joseph, D., Aribisala, A., Eshiett, E. P., Itopa, R., & Kpaku, G. (2015). Diagnostic reference level for adult brain computed tomography scans: A case study of a tertiary health care center in Nigeria. IOSR Journal of Dental and Medical Sciences, 14, 66–75.
Olowookere, C. J., Babalola, I. A., Jibiri, N. N., Obed, R. I., Bamidele, L., & Ajetumobi, E. O. (2012). A preliminary radiation dose audit in some Nigerian hospitals: Need for determination of national diagnostic reference levels (NDRLs). The Pacific Journal of Science and Technology, 13(1), 487–495.
Saravanakumar, A., Vaideki, K., Govindarajan, K. N., & Jayakumar, S. (2014). Establishment of diagnostic reference levels in computed tomography for select procedures in Pudhuchery, India. Journal of Medical Physics, 39, 50–55.
European Commission. (1997). Council Directive 97/43/Euratom of June 30, 1997 on health protection of individuals against the dangers of ionising radiation in relation to medical exposure. Official Journal of the European Communities, L180, 22–27.
Olarinoye, I. O., & Sharifat, I. (2010). A protocol for setting dose reference level for medical radiography in Nigeria: A review. Bayero Journal of Pure and Applied Sciences, 3, 138–141.
Elojeimy, S., Tipnis, S., & Huda, W. (2010). Relationship between radiographic techniques (kilovolt and milliampere-second) and CTDIvol. Radiation Protection Dosimetry, 141, 43–49.
Huda, W., & Mettler, F. A. (2011). Volume CT dose index and dose-length product displayed during CT: What good are they? Radiology, 258, 236–242.