X-RAY BARRIER ESTIMATION - A CASE STUDY OF THE GENERAL RADIOGRAPHY ROOM OF A MAJOR NIGERIAN TEACHING HOSPITAL

Main Article Content

Esien-Umo E.
Mallam S. P.
Akpa T. C.
Ukpong E. V.

Abstract

Background: The observation that the designing rules and calculations on shielding barrier thickness estimation for most general radiographic rooms are not being practised made the evaluation of the already existing shielding barrier of the General Radiography Room of Ahmadu Bello University Teaching Hospital, Shika - Zaria to ascertain its shielding adequacy very necessary.


Materials and Methods: The General Radiography Room with an area of 6.36m x 5m houses a Silhouette General X-ray machine; model MS-185N with serial number 0877 manufactured by General Electric (GE) Medical Systems in September, 2004. The evaluation was based on the method by NCRP Report No. 147, and used the number of patients examined in this room per week for shielding calculations.


Results: The results showed that the barrier thickness required varied from 0.10 to 0.46mm of lead for the secondary barriers and 1.36mm of lead for the primary barrier. The ratio of the calculated to the current (already contructed) barrier thickness ranged from 0.09 to 0.23 for the secondary barriers and 0.68 for the primary barrier .


Conclusion: Based on the results from this study, the shielding barrier already provided in the General Radiography Room of Ahmadu Bello University Teaching Hospital, Shika - Zaria is adequate and in the event of public access to the surrounding area, the dose rates are at acceptable limits.

Downloads

Download data is not yet available.

Article Details

Section

Articles

Author Biography

Esien-Umo E., Department of Radiography & Radiological Sciences, University of calabar

Radiography & Radiological Sciences Department, University of Calabar, Calabar

How to Cite

E., E.-U., P., M. S., C., A. T., & V., U. E. (2026). X-RAY BARRIER ESTIMATION - A CASE STUDY OF THE GENERAL RADIOGRAPHY ROOM OF A MAJOR NIGERIAN TEACHING HOSPITAL. Journal of Radiography and Radiation Sciences, 27(1), 16-24. https://doi.org/10.82547/jrrs.vol27no1.112

References

Sutton DG, Martin CJ, Williams JR, and Peet DJ. Radiation Shielding for Diagnostic Radiology. Second Edition, 2012. British Institute of Radiology, 36 Portland Place, London W1B 1AT, United Kingdom.

Casey M. Technical Note: Room Shielding Requirements in Diagnostic Radiology. British Journal of Radiology, 67 (1994).

National Council on Radiation Protection and Measurements (NCRP). Structural Shielding Design for Medical X-Ray Imaging Facilities. NCRP Report No. 147. Bethesda, MD: NCRP, 2004.

National Council on Radiation Protection and Measurements (NCRP). Evaluation for Medical Use of X-Rays and Gamma Rays of Energies up to 10 MeV. NCRP Report No. 49. Bethesda, MD: NCRP, 1976.

Archer BR. Recent History of the Shielding of Medical X-Ray Imaging Facilities. Health Physics, 2005; 88: 579–586.

Archer BR, and Gray JE. Important Changes in Medical X-Ray Imaging Facility Shielding Design Methodology: A Brief Summary of Recommendations in NCRP Report No. 147. Medical Physics, 2005; 32: 3599–3610.

Archer BR. Radiation Design Criteria for Diagnostic X-Ray Facilities. Proceedings of the Health Physics Society for Radiological Protection, 1993; 69–89.

Simpkin DJ, and Dixon RL. Secondary Shielding Barriers for Diagnostic X-Ray Facilities: Scatter and Leakage Revisited. Health Physics, 1998; 74(3): 350–365.

Dixon RL, and Simpkin DJ. Primary Shielding Barriers for Diagnostic X-Ray Facilities: A New Model. Health Physics, 1998; 74(2): 181–189.

Simpkin DJ. Shielding a Spectrum of Workload in Diagnostic Radiology. Health Physics, 1991; 61(2): 259–261.

National Council on Radiation Protection and Measurements (NCRP). Limitations of Exposure to Ionizing Radiation. NCRP Report No. 116. Bethesda, MD: NCRP, 1993.

The Health Physics and Radiological Health Handbook: Exposure and Shielding from External Radiation. Scinta Inc. Publishers, USA, 1993.

Turner JE. Atoms, Radiation, and Radiation Protection. Third Edition. Wiley-VCH Verlag GmbH & KGaA, 2007; 484 p.

Esien-Umo E, Mallam SP, Ishiekwen CA, and Egbe NO. Determination of Workload and Use Factors in a Major Nigerian Teaching Hospital. Nigerian Journal of Medical Imaging and Therapy, 2012; 20–24.