PATIENT X-RAY ENTRANCE SURFACE DOSE AT A TERTIARY HOSPITAL IN SOKOTO, NORTH-WEST NIGERIA
Main Article Content
Abstract
Background: The entrance surface dose (ESD) has been used to determine the diagnostic reference levels (DRL) by regulatory bodies to help medical radiation workers to checkmate excessive dose to patients.
Objective: To peer review practice in our centre using recommended diagnostic reference levels (DRLs) as standard.
Methods: A prospective cross-sectional study of fifty x-ray patients who had thermoluminescent dosimeters (TLD-100) taped on their skin to quantify entrance surface dose (ESD) during x-ray investigation. The patients appeared in Radiology Department of Usmanu Danfodiyo University Teaching Hospital (UDUTH), Sokoto between 1st August and 30th September 2018. Their weights and heights were measured directly. Thereafter, they were divided into 5 groups of ten each for five specific anatomical areas: PA chest, AP of pelvis, abdomen, lumbosacral spine, and lateral lumbosacral spine. Irradiated TLDs were read, and the mean ESD was calculated.
Results: Entrance surface dose (ESD) was 0.26 mGy (chest), 0.46 mGy (AP lumbosacral), 0.71 mGy (pelvis), 1.3 mGy (abdomen), and 1.6 mGy (lateral lumbosacral). The ESD values were found to be lower than similar studies carried out in Nigeria, as well as diagnostic reference levels in UK.
Conclusion: Optimization of patient protection in UDUTH was comparable to recommended practices locally and internationally.
Downloads
Article Details
Section

This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
All articles in JRRS are published under the Creative Commons Attribution 4.0 International License (CC BY 4.0). This permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
How to Cite
References
Egbe NO, Inyang SO, Ibeagwa OB, Chiaghanam NO. Pediatric radiography entrance doses for some routine procedures in three hospitals within eastern Nigeria. J Med Phys, 2008; 33(1): 29 – 34.
ICRP. Recommendations of the International Commission on Radiological Protection. ICRP publication 60. Ann ICRP 1991; 21:1-3.
ICRP. Radiological protection and safety in medicine. ICRP publication 73. Ann ICRP 1996; 26:1-47.
Nigerian Nuclear Regulatory Authority. Nigerian Radiation Safety Regulation for Diagnostic and Interventional Radiology Practice. 2006; 19: 1-33.
Joseph DZ, Chinedu O, Nkubli FB, Luntsi G, Shem SL, Dlama Y. Rationale for Implementing Dose Reference Level as a Quality Assurance Tool in Medical Radiography in Nigeria. IOSR Journal of Dental and Medical Sciences, 2014; 13(12): 41-45.
Joseph DZ, Nzotta CC, Skam JD, Umar MS, Dambele MY. Diagnostic reference levels for mammography examinations in North Eastern Nigeria. African Journal of Medical and Health Sciences, 2018; 17(1): 54 – 59.
Akpochafor MO, Omojola AD, Adeneye SO, Aweda MA, Ajayi HB. Determination of reference dose levels among selected x‑ray centers in Lagos State, South‑West Nigeria. J Clin Sci 2016; 13:167‑172.
Ekpo E, Adejoh T, Akwo J, Onwujekwe EC, Ali AA, Abba M, Adesina K, Omiyi D, Chiegwu HC. Diagnostic reference levels for common computed tomography (CT) examinations: results from the first Nigerian nationwide dose survey. Journal of Radiological Protection, 2018; 38:525 – 535.
Ezenma I, Isani B, Moi SA. Comparative study of patient entrance dose of two diagnostic centres in Ekiti state, Nigeria with diagnostic reference levels. Nigeria Journal of Medical Imaging and Radiation Therapy. 2013; 2(1): 23 – 29.
Nzotta CC, Egbe NO, Adejoh T, Nkubli FB, Ezeador IS. Radiation Dose from Exposure to Computed Tomography Scan of the Brain in a Reference Hospital in Nigeria. Pakistan Journal of Radiology, 2016; 26(2): 77-80.
Njiti MM, Nwobi IC, Garba I, Ahidjo A, Bashir T, Abubakar A, Hamidu AU, Luntsi G. Estimation of entrance surface dose to adult patients undergoing plain chest radiographic examinations in a Northern Nigeria population. Kanem Journal of Medical Sciences. 2014; 8(1): 27-32.
European Commission (EC). European guidelines on quality criteria for diagnostic radiographic images. Brussels, Luxembourg: EUR 16260-EN; 1996. Available at https://www.sprmn.pt/pdf/EuropeanGuidelineseur16260.pdf
[Accessed on 29/5/2019].
Ishiekwen AC, Aliyu AS, Nwafor MS. X-Ray Dose Audit and Potential Local Diagnostic Reference Levels in Select Hospitals in Kaduna State, Nigeria. J Rad & Radiat Sci, 2018; 32(1):104 – 109.
Jumaa YT, Mohamed Y, Khadija M, Ahmed A, Abdelmoneim S. Assessment of entrance surface dose for patients from common radiology examinations in Sudan. Life Sci J, 2014; 11(2):164 -168.
Gholami M, Maziar A, Khosravi HR, Ebrahimzadeh F, Mayahi S. Diagnostic reference levels for routine x-ray examinations in Lorestan Province, Iran. International Journal of Radiation Research, 2015; 13(1): 85-90.
Hart D, Hillier MC, Shrimpton PC. Doses to patients from radiographic and fluoroscopic x-ray imaging procedures in the U.K: 2010 review. Report number: HPA-CRCE–034, 2012; 1-95. Available at https://assets.publishing.service.gov.uk/government/uploads/system/uploads/attachment_data/file/431134/HPA-RPD-029.pdf
[Accessed on 29/5/2019].