ASSESSMENT OF RADIATION SAFETY MEASURES IN SELECT RADIO-DIAGNOSTIC CENTRES IN KADUNA STATE, NIGERIA
Main Article Content
Abstract
Background: The risks associated with x-rays in radiography are minimized through international best practices of a regular program of quality control.
Objective: To assess the radiation safety precautions taken in five radio-diagnostic centers in Kaduna State, Nigeria.
Methodology: Measuring tapes, radiation meters and other accessories were used to assess for x-ray room dimension, x-ray tube leakage, kVp accuracy, half value layer (HVL), mA linearity, optical and x-ray beam congruence, beam alignment, and timer accuracy.
Results: Only one centre complied with the 16m² minimum room dimension required for an x-ray room, passed area monitoring without a compromise, and wholly complied with kVp error limit of 5%. Two centres passed optical radiation beam congruence, beam alignment and timer accuracy tests. Three centres passed mA linearity test. All the five centers passed the tube leakage test with none of the centers recording up to 1 mSv/h at 100 cm from the tube surface.
Conclusion: X-ray machines used in Kaduna States are safe. However, there are uncoordinated attention to other safety precautions. More efforts should be made to ensure that holistic regulatory standards are met in order to consolidate radiation protection.
Downloads
Article Details
Section

This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
All articles in JRRS are published under the Creative Commons Attribution 4.0 International License (CC BY 4.0). This permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
How to Cite
References
European Commission. (2000). European guidelines for quality criteria for computed tomography (EUR 16262 EN). Luxembourg: European Commission.
Mahesh, M. (2002). Search for isotropic resolution in CT from conventional through multiple-row detector. RadioGraphics, 22, 949–962.
International Commission on Radiological Protection. (2009). The history of ICRP and the evolution of its policies (ICRP Publication 109). Oxford: Elsevier Science Ltd.
Vano, E., & Fernandez Soto, J. M. (2007). Patient dose management in digital radiography. Biomedical Imaging and Intervention Journal, 3(2), e26.
Taha, M. T., Al-Ghorabie, F. H., Kutbi, R. A., & Saib, W. K. (2015). Assessment of entrance skin doses for patients undergoing diagnostic X-ray examinations in King Abdullah Medical City, Makkah, KSA. Journal of Radiation Research and Applied Sciences, 8(1), 100–103.
Adejumo, S. B., Irurhe, N. K., Olowoyeye, A. O., Ibitoye, A. Z., Eze, C. U., & Omiyi, O. D. (2012). Evaluation of compliance to radiation safety standards among radiographers in radiodiagnostic centres in South West Nigeria. World Journal of Medical Sciences, 7(3), 194–196.
Mettler, F. A. (2012). Medical effects and risks of exposure to ionizing radiation. Journal of Radiological Protection, 32(9), N9–N13.
Aweda, M. A., & Awosanya, G. O. (2007). Safety awareness in selected X-ray diagnostic centres in Surulere and Ikeja Local Government Areas of Lagos State, Nigeria. Nigerian Journal of Health and Biomedical Sciences, 6(1), 62–72.
Geeta, S. P. (2005). Medical equipment in government health facilities: Missed opportunities. Indian Journal of Medical Sciences, 59(1), 13–19.
Sungita, Y. Y., Simon, S. L., & Peter, M. (2006). Diagnostic X-ray facilities as per quality control performance in Tanzania. Journal of Applied Clinical Medical Physics, 7(4), 66–73.
Okeji, M. C. (2010). Radiation exposure from diagnostic radiography: An assessment of X-ray beam collimation practice in some Nigerian hospitals. Internet Journal of Medical Update, 5(2), 31–33.
Eze, K. C., Nzotta, C. C., Marchie, T. T., Okegbunam, B., & Eze, T. E. (2011). The state of occupational radiation protection and monitoring in public and private X-ray facilities in Edo State, Nigeria. Nigerian Journal of Clinical Practice, 14(3), 308–310.
International Atomic Energy Agency. (2004). Optimization of protection in radiography: IAEA training material on radiation protection. Vienna, Austria: IAEA.
Fung, K. K., & Gilboy, W. B. (2001). The effect of beam tube potential variation on gonad dose to patients during chest radiography using TLD. British Journal of Radiology, 74(880), 358–367.
Rehani, M. M. (1995). Diagnostic imaging quality assurance (pp. 44–106). India: Jaypee Brothers Medical Publishers Ltd.